匠析生物科技

咨询热线

86 18118137745

Agilent 7697A顶空进样器的进样针堵塞问题该怎么维修清理?

0 浏览 0 赞同 2026-03-11

For the Agilent 7697A headspace sampler's injection needle blockage issue, you can refer to the following solutions: Possible causes: 1. Particulate matter or high boiling point impurities in the sample deposit inside the needle; 2. Residual substances solidify due to long-term non-use. Recommended troubleshooting steps:

  1. Cool the GC inlet to a safe handling temperature and the GC oven to room temperature to avoid column damage, then turn off the carrier gas.
  2. Use an appropriate solvent to repeatedly flush the injection needle to remove blockages.

If you encounter difficulties in handling this issue or need professional maintenance services, 匠析生物 can provide comprehensive solutions. Our main business includes one-stop services for laboratory instruments, consumables, spare parts, and after-sales repair and maintenance, covering brands such as Agilent, Thermofisher, Waters, etc. For Agilent 7697A headspace samplers, we offer professional repair, cleaning, and maintenance services, including needle blockage resolution, spare part replacement if needed. Additionally, we provide instrument leasing, dust cleaning, chromatographic application consulting, and other services to meet your diverse laboratory needs.

匠析生物

匠析生物

色谱分析专家 · 高级工程师

更多仪器使用和维护问题,欢迎电话咨询:

+86 181 1813 7745