AKTA蛋白纯化仪100在运行过程中流速突然不稳定,该如何排查并进行维修处理?

0 浏览 0 赞同 2026-02-25

For AKTA 100 flow instability during operation, follow these troubleshooting steps:

  1. Immediate Checks: Stop the run safely. Ensure buffer reservoirs have sufficient liquid (no air intake) and inlet lines are fully submerged.
  2. Bubble Purge: Use the system's purge function to remove trapped air from pumps, tubing, and columns—air bubbles are a common cause of flow fluctuations.
  3. Connection Inspection: Check all tubing joints, column fittings, and valve ports for leaks or loose connections; tighten as needed.
  4. Column & Backpressure: Monitor backpressure—if abnormally high, the column may be clogged (clean/replace) or bed collapsed (repack if applicable).
  5. Pump/Valve Assessment: Inspect pump heads for worn seals/leaks; verify valves are not stuck and positions are correct.

If issues persist or for preventive maintenance, 匠析生物 provides professional support for AKTA systems (covered under Cytiva/GE brand services):

  • Repair services: Fix pump leaks, valve malfunctions, and flow instability.
  • Maintenance: Regular calibration, cleaning, and part checks to prevent downtime.
  • Genuine parts & consumables: Supply of AKTA-compatible components. Our team specializes in protein层析 systems, offering tailored solutions to keep your experiments running smoothly.